Loop correction | Hardware | Correction of floating capacitance between collector supply and ground |
| Software | Pseudo-looping through software optimization |
Step generator | Offset | Setting range | STEP AMPLITUDESetting -10x to +10x |
| Resolution | STEP AMPLITUDE1% of setting |
Current mode | Range | 50nA to 200mA per step, 21 steps 1-2-5 switching |
| Maximum current | 2A |
| Maximum voltage | 10V or more |
Voltage mode | Range | 50mV to 2V per step, 6 steps 1-2-5 switching |
| Maximum current | ±40V |
| Maximum voltage | 500mA~(~8V), 200mA~(~15V), 10mA~(~40V) |
| Step rate | 2x 50Hz or 60Hz (1x in AC mode), pulse interval in HC mode |
Pulse step | Pulse width | 50μs~400μs(10μs step) |
| When in HC mode, the pulse width is 100μs wide before and after the pulse of the collector supply |
| Number of steps | 0~20Step |
| AUX output | Range | OFF, -40V~+40V(Variable in 100mV steps) |
| Measurement mode | REPEAT, STOP/SINGLE, SWEEP |
Vertical axis (full scale 10div) | Collector current | Range | HV mode: 1μA/div to 2A/div, 20 steps 1-2-5 switching (HC mode is not included) |
Accuracy | 2% + 0.05 x VERT/div of readout |
| To this, add the loop correction error for the following maximum peak voltages |
| 0.5μA(30V), 1μA(300V), 6μA(3kV), 12μA(5kV) |
| Specify at least 10% of the maximum peak voltage for 30V, 300V, and 3kV, and at least 30% for 5kV |
| Emitter current | Range | 1nA/div~2mA/div,20 step 1-2-5 switching |
| (LEAKAGE) | (Collector supply mode is LEAKAGE) |
| Accuracy | 2% + 0.05 x VERT/div + 1nA or less of readout |
Horizontal axis (full scale 10div) | Collector voltage | Range | HC mode: 50mV/div to 5V/div, 7 steps 1-2-5 switching (HV mode is not included) |
| Accuracy | 2% + 0.05 x HORIZ/div or less of readout |
Base/emitter voltage | Range | 50mV/div~5V/div,7 step 1-2-5 switching |
| Accuracy | 2% + 0.05 x HORIZ/div or less of readout |
Display | Display | 8.4-inch TFT LCD |
Number of data points | 1,000 points/trace (AC, full wave rectification) |
| 20-1,000 points/trace (SWEEP mode) |
| Trace display | Point interpolation display, dot display |
| Averaging | OFF,2 to 255 times |
| Persistence | OFF, SHORT, LONG,Infinite length |
| Save internal waveform(REF) | 4-Panel |
Cursor | DOT | Vert, Horiz, β or gm |
| fLINE | Vert, Horiz, 1/grad, intercept |
| FREE | Vert, Horiz, β or gm |
| WINDOW | Vert, Horiz, β or gm in the WINDOW area |
Data save/read | Memory | Setup: 256 pcs, REF: 4 screens |
| External memory | USB1.1: Setup, waveform, screen hardcopy (BMP, TIFF, PNG) |
| Remote | Equipped with remote 10BASE-T/100BASE-TX 1 port via LAN connection |
| Power supply | CS-3xxx | AC100-240V 50/60Hz,Maximum operating power:500VAStandby:7W Max |
| External dimensions(mm) | CS-3100 | Approx. 424W x 220H x 555D (excluding protrusions and accessories), approx. 28kg (excluding accessories) |
| Weight | CS-3200,3300 | Approx. 424W x 354H x 555D (excluding protrusions and accessories), approx. 43kg (excluding accessories) |